Abstract
Purpose: To determine the correlation between fundus perimetry with Micro Perimeter 1 (MP1) and conventional automated static threshold perimetry using the Humphrey Field Analyzer (HFA) in healthy individuals and in subjects with glaucoma. Materials and methods: In this study, we enrolled 45 eyes with glaucoma and 21 eyes of age-matched, healthy individuals. All subjects underwent complete ophthalmic examination. Differential light sensitivity was measured at 21 corresponding points in a rectangular test grid in both MP1 and HFA. Similar examination settings were used with Goldmann III stimulus, stimulus presentation time of 200 ms, and white background illumination (1.27 cd/m(2)). Statistical analysis was done with the SPSS 14 using linear regression and independent t-test. Results: The mean light thresholds of 21 matching points in control group with MP1 and HFA were 14.97 ± 2.64 dB and 30.90 ± 2.08 dB, respectively. In subjects with glaucoma, the mean values were MP1: 11.73 ± 4.36 dB and HFA: 27.96 ± 5.41 dB. Mean difference of light thresholds among the two instruments was 15.86 ± 3.25 dB in normal subjects (P < 0.001) and 16.22 ± 2.77 dB in glaucoma subjects (P < 0.001). Pearson correlation analysis of the HFA and MP1 results for each test point location in both cases and control subjects showed significant positive correlation (controls, r = 0.439, P = 0.047; glaucoma subjects, r = 0.812, P < 0.001). There was no difference between nasal and temporal points but a slight vertical asymmetry was observed with MP1. Conclusion: There are significant and reproducible differences in the differential light threshold in MP1 and HFA in both normal and glaucoma subjects. We found a correction factor of 17.271 for comparison of MP1 with HFA. MP1 appeared to be more sensitive in predicting loss in glaucoma.
Keywords
Publication details
- DOI
- 10.4103/0974-620x.99372
- Journal
- Oman Journal of Ophthalmology, 5(2), 97-102
- Publisher
- Medknow Publications
- Open access
- Gold open access
Cite this article
APA 7
Ratra, V., Ratra, D., Gupta, M., & Vaitheeswaran, K. (2012). Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects. Oman Journal of Ophthalmology, 5(2), 97-102. https://doi.org/10.4103/0974-620x.99372
MLA 9
Ratra, Vineet, et al. "Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects." Oman Journal of Ophthalmology, vol. 5, no. 2, 2012, pp. 97-102. https://doi.org/10.4103/0974-620x.99372.
Chicago (author–date)
Ratra, Vineet, Dhanashree Ratra, Muneeswar Gupta, and K Vaitheeswaran. 2012. "Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects." Oman Journal of Ophthalmology 5 (2): 97-102. https://doi.org/10.4103/0974-620x.99372.
Harvard
Ratra, V., Ratra, D., Gupta, M. and Vaitheeswaran, K. (2012) 'Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects', Oman Journal of Ophthalmology, 5(2), pp. 97-102. doi:10.4103/0974-620x.99372.
Vancouver
Ratra V, Ratra D, Gupta M, Vaitheeswaran K. Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects. Oman Journal of Ophthalmology. 2012;5(2):97-102. doi:10.4103/0974-620x.99372
IEEE
V. Ratra, D. Ratra, M. Gupta, and K. Vaitheeswaran, "Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects," Oman Journal of Ophthalmology, vol. 5, no. 2, pp. 97-102, 2012, doi: 10.4103/0974-620x.99372.